Description

Flow-through open amperometric cell 6x8 mod. ECL6 - EMEC
For the measurement of free chlorine (organic and inorganic)

The open amperometric cells consist of a flow electrode holder with a chlorine sensing electrode and a flow sensor.
Some models can house up to a maximum of three additional probes (temperature, pH and Redox).
The flow rate of water flowing inside this cell must be kept constant and at a rate of 40 l/h.
A pressure stabilizer is available for applications in places subject to sudden changes in pressure.
It is recommended to install a filter before the probe holder.

Principle of operation
ECL probes have an open cell amperometric measurement system. The measurement method is an electrochemical technique that measures the current developed between two electrodes (platinum and copper or platinum and silver).

Characteristics:
- Parameter: free chlorine (organic and inorganic)
- Measurement range: 0-10 mg/l (0-10 ppm) resolution: ± 0.05
- Connection: 2 wires (+red; -black)
- Measurement system: amperometric with 2 electrodes (platinum/copper; gold/copper on request)
- pH range: 6-8 pH
- Execution time: first polarization: about 2 hours; re-polarization: 50 min. about
- Response time: T90: 2 min. about
- Calibration: ref. Operating manual: "Calibration"
- Alkalinity: 100ppm
- Working temperature: 5-40°C (41-104°F)
- Pressure: 0.4 - 5 bar (5.8 - 72.5 PSI)
- Cable (standard): 2 m (6.6 ft); 1m if panel mounted
- Workflow: 40L/h
- Probe holder for: pH, ORP and temperature
- Fittings for connection to the sample line: 6x8
- Material: platinum/copper electrode; measuring cell in PPMA (methacrylate)
- Mounting: on vertical flat surface (panel, stand, etc.)
- Storage: protect from frost, in a cool and dry place (5-40° C)
- Maintenance: Regular signal check. SHORTEN THE MAINTENANCE INTERVALS ACCORDING TO THE QUALITY OF THE WATER.